<img alt="" src="/images/1/f/e/4/a/1fe4aad118f5911a24db39c0267ac96175490e2d-vakuum-mil-std-286c.jpg?name" />

Vacuum Stability Test

Determining the thermal stability of nitro-cellulose and other nitro-compounds.

MIL-STD-286C STANAG 4556

Apparatus for determining the thermal stability of nitro-cellulose and other nitro-compounds, single and double base propellants according to the MIL-STD -286C method 403.1.3 and STANAG 4556, Procedure 1, Vacuum Stability Test (by manometer method).


FEATURES

  • Heavy aluminium heating block, electrical heated; 15 holes 15.75 mm Ø, depth 125 mm; temperature sensor: PT100 probe; excess temperature safety device 150°C
  • Programmable heating controller (PID) in a separate control box, length of the power cable: appr. 2 m; digital display for actual and set temperature; control accuracy: ±0.1°C; max. adjustable temperature: 135°C
  • Set of 15 test tubes consisting of the cylindrical test tube with mercury collar, capillary tube with mercury cup; dimensions of test tube: Ø 15 mm x 155 mm
  • Rack for filling glass assembly with mercury
  • Vacuum pump IP44: 4 stage diaphragm pump, max. vacuum: 0.6 mbar, max. suction capacity: 13 l/min

Technical Specification

Mains:230 V / 50 – 60 Hz
Power:1800 VA